7.11Optimizing Your Process System with the WATLOW Series 988
How to Choose the Right 988 to Fit Your Application
Series 988/989 Model Number Information
98 _ _ - _ _ _ _ - _ _ _ _
Hardware
6 = Vertical mounting, 24-28 V
7 = Horizontal mounting, 24-28 V
8 = Vertical mounting, 100-240 V
9 = Horizontal mounting, 100-240 V
Software
A = Standard
B = Enhanced (with cascade, ratio, dual PID, differential)
#1 Input
1 = Basic thermocouple signal conditioner
(excluding Type R, S and B)
2 = Universal signal conditioner
#2 Input
0 = None
1 = Basic thermocouple signal conditioner
(excluding Type R, S and B)
2 = Universal signal conditioner
3 = Slidewire feedback
4 = Current transformer
5 = Second digital event (one digital event is standard on all units)
#1 Output
B = Solid state relay, Form A, 0.5A, with RC suppression
C = Switched DC, open collector, isolated
D = *Electromechanical relay, Form C, 5A, with RC supression
E = *Electromechanical relay, Form C, 5A, without RC suppression
F = Universal process, 0-5VDC, 1-5VDC, 0-10VDC, 0-20mA, 4-20mA
K = Solid state relay, Form A, 0.5A, without RC suppression
#2 Output
A = None
B = Solid state relay, Form A, 0.5A, with RC suppression
C = Switched DC, open collector, isolated
D = *Electromechanical relay, Form C, 5A, with RC suppression
E = *Electromechanical relay, Form C, 5A, without RC suppression
F = Universal process, 0-5VDC, 1-5VDC, 0-10VDC, 0-20mA, 4-20mA
K = Solid state relay, Form A, 0.5A, without RC suppression
T = External signal conditioner power supply, 5, 12 or 20VDC @ 30mA
#3 Output
A = None
B = Solid state relay, Form A, 0.5A, with RC suppression
C = Switched DC, open collector, isolated
J = *Electromechanical relay, Form A or B, 5A, without RC suppression
K = Solid state relay, Form A, 0.5A, without RC suppression
M = Retransmit, 0-20mA, 4-20mA
N = Retransmit, 1-5VDC, 0-10VDC, 0-5VDC
T = External signal conditioner power supply, 5, 12 or 20VDC @ 30mA
#4 Output
A = None
B = Solid state relay, Form A, 0.5A, with RC suppression
C = Switched DC, open collector, isolated
D = *Electromechanical relay, Form C, 5A, with RC suppression
E = *Electromechanical relay, Form C, 5A, without RC suppression
K = Solid state relay, Form A, 0.5A, without RC suppression
R = Isolated RS-232 communications
S = Isolated EIA-485/RS-422 communications
T = External signal conditioner power supply, 5, 12 or 20VDC @ 30mA
Display Color (Upper/Lower)
GG = Green/Green
GR = Green/Red
RG = Red/Green
RR = Red/Red
Sensor Ranges
J t/c: 32 to 1500°F
0 to 816°C
K t/c: -328 to 2500°F
-200 to 1371°C
T t/c: -328 to 750°F
-200 to 399°C
N t/c: 32 to 2372°F
0 to 1300°C
R t/c: 32 to 3200°F
0 to 1760°C
S t/c: 32 to 3200°F
0 to 1760°C
B t/c: 1598 to 3300°F
870 to 1816°C
E t/c: -328 to 1470°F
-200 to 799°C
C t/c: 32 to 4200°F
0 to 2316°C
D t/c: 32 to 4200°F
0 to 2316°C
Pt 2: 32 to 2543°F
0 to 1395°C
1°RTD (JIS): -328 to 1166°F
-200 to 630°C
1°RTD (DIN): -328 to 1472°F
-200 to 800°C
0.1°RTD (JIS
and DIN): -99.9 to 999.9°F
-73.3 to 537.7°C
0-5VDC: -999 to 9999
1-5VDC: -999 to 9999
0-10VDC: -999 to 9999
0-20mA: -999 to 9999
4-20mA: -999 to 9999
0-50mVDC: -999 to 9999
0-100mVDC: -999 to 9999
Slidewire: 100 to 1200Ω
Current: 0 to 50mA
Potentiometer: 0 to 1200Ω
*Electromechanical relays are not recom-
mended for PID control. They are warranted
for only 100,000 contact closures
100Ω platinum
Commentaires sur ces manuels